1、手持式合金分析儀操作規(guī)程(中英文)一、范圍1. Range本規(guī)程適用于廠內(nèi)型號為 Niton XLT898W 手持式合金分析儀的使用操作。This operating regulation is applie for hany-alloy analysis in HBG name Niton XLT898W.二、工作原理及適用范圍:2. Working principle an scope of application采用 X 射線熒光( XRF )技術(shù)進行多元素分析。用于各種高低合金鋼、不銹鋼、工具鋼、鉻 / 鉬鋼、鎳合金、鈷合金、鎳 / 鈷耐熱合金、鈦合金、銅合金等,可分析 Ti V Cr

2、 Mn Fe Co Ni Cu Zn Nb Zr Mo Ag P Sn Hf Ta W Re Pb Bi Se Sb 等元素。Using fluorescent X-ray technology to have multi-element analysis. this evie is applie for kins of alloye steel, non-corrosive steel, tool steel, chrome-molybenum steel, nickel alloy, cobalt alloy, nickel/cobalt heat-resisting alloy, tit

3、anium alloy, copper alloy an so on. it can also analysis kins of elements like Ti V Cr Mn Fe Co Ni Cu Zn Nb Zr Mo Ag P Sn Hf Ta W Re Pb Bi Se Sb.三、工作條件:3.Working conition 工作溫度: -20 -50 ;樣本溫度: 0 -50 。Working temperature:-20 -50 ;specimen temperature: 0 -50 .四、被測金屬樣本的外形尺寸:4.outline imension of measure

4、 metal specimen線材:直徑 1mm 以上;小口徑管材:外徑 2mm 以上;被測材料與探測窗之間的距離小于 10mm ;被測材料(表面)的狀態(tài):平面、曲面、粗糙表面、不規(guī)則表面、粉末狀、振動狀態(tài)等。被測表面應打磨露出金屬光澤,不影響分析結(jié)果。Wire stock: iameter greater than 1mm ;Small pipe-tube: external iameter greater than 1mm ;istance between measure material an etecting winow less than 10mm ;Status of measur

5、e material(surface):flat face, curve face, rough face, irregular face, pulverous face, vibrate face an so on, measure material surface shoul be shine until the material appear metal brightness an not effect on analysis result.五、操作步驟:5.Operation proceure:1.使用前認真閱讀 本 設備操作使用規(guī)程 或使用說明書 ,仔細檢查儀器表面,如發(fā)現(xiàn)有鮮明破損

6、或異常現(xiàn)象應立即更換。2.使用該設備時應正確佩戴設備防護腰帶,手握設備時應先把防滑帶套入手腕,以防設備損壞。3.測量前應保證被測面露出金屬光澤。按下電源開關(guān),儀器發(fā)出提示音,液晶屏幕顯示數(shù)據(jù)后,直接在觸摸屏上輸入 1 、 2 、 3 、 4 、 E 進入操作界面(主界面) 。4.儀器校正:在正式測試前,應檢測“標塊”,確認檢測值正常,方可進檢測產(chǎn)品。5.測量時把發(fā)射鏡貼近被測物體表面,按下把手上的檢測開關(guān)按鈕,此時工作提醒燈閃爍測量開始,待儀器發(fā)出提示音測量結(jié)束。6.按“” / “”鍵翻看檢測數(shù)據(jù)。繼續(xù)檢測請則再次按下把手上的檢測開關(guān)按鈕。退出檢測界面按“”回到主界面。7.測量完畢后長按電源開

7、關(guān),關(guān)機,并且將設備放置在防水設備箱內(nèi)。1. Reaing the operation stanar or operation instruction of this evice carefully before using, examining the evice surface carefully, Replacing the evice immeiately if any abnormal phenomenon or breakage was foun.2. Wearing evice proofing girle before using, antiski belt shoul be

8、grabbe in wrist when grab the evice in case of amaging.3. Metal brightness shoul be appeare before measurement. Pressing the battery button, then the evice come out a present soun. After the liqui crystal screen manifests recor, press 1 2 3 4 E in the touch screen to get in operation interface.4. ev

9、ice ajustment: before formal testing, testing block shoul be teste, the prouct shoul not be testing until the recor was normal.5. The etective winow shoul near the measure prouct surface when the measurement is begin, press the etecting button in grip, when working light is flashing the measurement

10、begin, it is over when the evice comes out soun.6. Pressing “” / “” to observe etection recor. Pressing the etecting button again if you want to continue etection. Pressing “” to quit etection interface.7.Holing on pressing power button when measurement was complete, an the evice shoul be locate in

11、water-proof instrument case.六、注意事項:6. matters neee attention:1.本儀器需專人操作,專人負責日常維持、清潔及保養(yǎng)工作,并且做好儀器使用記錄。非相關(guān)人員不得擅自操作。2.操作前應確保電池有足夠電量。3.在任何時候都要確定 X 射線方向、謹防輻射。4.在任何情況,如果未使用儀器,不得開啟擋板。5.在測試進程中不得將儀器放在 Kapton 窗口附近。在擋板開啟時,勿將儀器指向自己或其他任何人。6.在下載讀數(shù)時不可進行測量,在刪除數(shù)據(jù)時切勿關(guān)閉儀器。7.勿將儀器侵入水中。8.本儀器因采用 X 射線熒光 , 因而要避免在露天強光(特別是紫外線強

12、的前提下)操作,以免數(shù)據(jù)不準。9.本儀器為精密設備,操作時盡快做到輕拿輕放,以免設備破損。1.This evice nee specialize operator an specialize iniviual to maintain, cleaning aily, evice recor shoul be written own an other people without permission shoul not use this evice.2. Ensure enough battery before operation.3. Ensure the erection of X-ray

13、whenever using it. Being aware of raiation.4. Uner any circumstance, baffle must not be opene if the evice was not use.5. The evice shoul not locate near the Kapton winow uring the testing phase. When baffle is open, the evice shoul not opoint to youself or others.6.No testing when the recor is ownl

14、oaing an No shutting own the evice when eleting the recor.7. The evice shoul not immerse into water. 8. Own to the evice using X-ray, the evice shoul avoi out-oor operation(especially in the ay when ultraviolet ray is high),in case of inaccuracy recor.9. This evice is a precise instrument, when operating it shoul place s

溫馨提示

  • 1. 本站所有資源如無特殊說明,都需要本地電腦安裝OFFICE2007和PDF閱讀器。圖紙軟件為CAD,CAXA,PROE,UG,SolidWorks等.壓縮文件請下載最新的WinRAR軟件解壓。
  • 2. 本站的文檔不包含任何第三方提供的附件圖紙等,如果需要附件,請聯(lián)系上傳者。文件的所有權(quán)益歸上傳用戶所有。
  • 3. 本站RAR壓縮包中若帶圖紙,網(wǎng)頁內(nèi)容里面會有圖紙預覽,若沒有圖紙預覽就沒有圖紙。
  • 4. 未經(jīng)權(quán)益所有人同意不得將文件中的內(nèi)容挪作商業(yè)或盈利用途。
  • 5. 人人文庫網(wǎng)僅提供信息存儲空間,僅對用戶上傳內(nèi)容的表現(xiàn)方式做保護處理,對用戶上傳分享的文檔內(nèi)容本身不做任何修改或編輯,并不能對任何下載內(nèi)容負責。
  • 6. 下載文件中如有侵權(quán)或不適當內(nèi)容,請與我們聯(lián)系,我們立即糾正。
  • 7. 本站不保證下載資源的準確性、安全性和完整性, 同時也不承擔用戶因使用這些下載資源對自己和他人造成任何形式的傷害或損失。

評論

0/150

提交評論